Function Checking and Testing 5-5
5.4.1
Check Flow
Check each imaging
mode
Image & video manage
Record and save the
exam
Check the measurements
Check in cine mode
5.4.2
Checking Content
5.4.2.1
B Mode
In B Mode scanning, the image parameter area on the corner of the screen will
display the real-time parameter values as follows:
Parameters F
D
G
FR
DR
Meaning
Frequency Depth Gain Frame Rate B Dynamic Range
Parameters that can be adjusted to optimize the B Mode image are indicated in the
following.
Adjustment
Items
Control Panel
Gain, Depth, TGC, iTouch, Focus, Frequency
Menu
Colorize, Colorize Map Gray Map, Line Density, U/D Flip,
Persistence, Smooth, Rotate, L/R Flip, Focus Number, FOV
Position, Frequency, Focus Position, iClear, Dyn Range, FOV,
TSI, Acoustic Power, Auto Merge, ExFov, Steer, iBeam, Hscale,
Gray Invert.