11-2 Troubleshooting of Regular Malfunctions
up normally
3
The AC input is normal ,AC power
indicator remains off;
Power indicators: off
AC-DC
power
board fails
Replace
power
module
4
AC power indicator: ON;
Power
–on status indicator: off after
power button pressed
Power indicators: off
AC_DC power
board: function
normal;
DC-DC power
board: fails
Replace
power
module
5
AC power indicator: ON;
Power
–on status indicator: blinks after
power button pressed
Power indicators: off
CPU module: fails
Replace
CPU
module
6
AC power indicator: ON;
Power on/off indicator: blinks and turns
on after pressing power button
Power indicator: 5V off or 3.3V off
DC-DC power
board: fails
Replace
power
module
11.2 System Cannot Start up Normally
11.2.1 Module or Board Related
No.
Descriptions
Remarks
1
CPU module
2
HDD
11.2.2 Key Points Supporting Troubleshooting
No.
Key Points Supporting Troubleshooting
Remarks
1
Character and progress status during the
starting of the system
2
Alarm and prompts during the starting of the
system
3
Backlight status during power on the control
panel
Support
to
check
if
the
malfunction occurs in power
supply module or CPU module
Содержание DC-25
Страница 1: ...DC 32 DC 30 DC 28 DC 26 DC 25 Digital Ultrasound System Service Manual Revision 11 0 ...
Страница 2: ......
Страница 7: ...v Appendix C Requirements of Performance Indices C 1 Appendix D Boot Screen D 1 ...
Страница 8: ......
Страница 12: ......
Страница 70: ......
Страница 88: ...5 18 Function Checking and Testing ...
Страница 94: ...6 6 Software installation and Maintenance ...
Страница 168: ......
Страница 178: ......
Страница 190: ......
Страница 206: ......
Страница 211: ...Boot Screen D 1 Appendix D Boot Screen BIOS Boot Screen LINUX Boot Screen ...
Страница 212: ...D 2 Boot Screen Doppler Boot Screen ...
Страница 213: ...P N 046 009019 00 11 0 ...