7-38 Structure and Assembly/Disassembly
Fig 7-33 Disassemble the front cover
2. Remove the M4x8 screws (16 pcs) which are used to secure the probe board
module, hold the two handles and stably pull out the probe board module.
Fig 7-34 Disassemble the Probe Board Assembly
7.4.8
Left Side-Panel Assembly
Remove the cross panhead screws (M4X8 X 2) fixing the mounting bracket to the left
side-panel. Remove the cross panhead screws (M4X8 X 1) fixing the back mounting ear
to the left side-panel. Remove the cross panhead screw (M4X8 X 1) fixing the front
mounting ear to the left side-panel. Then you can take out the left side-panel.
Содержание DC-25
Страница 1: ...DC 32 DC 30 DC 28 DC 26 DC 25 Digital Ultrasound System Service Manual Revision 11 0 ...
Страница 2: ......
Страница 7: ...v Appendix C Requirements of Performance Indices C 1 Appendix D Boot Screen D 1 ...
Страница 8: ......
Страница 12: ......
Страница 70: ......
Страница 88: ...5 18 Function Checking and Testing ...
Страница 94: ...6 6 Software installation and Maintenance ...
Страница 168: ......
Страница 178: ......
Страница 190: ......
Страница 206: ......
Страница 211: ...Boot Screen D 1 Appendix D Boot Screen BIOS Boot Screen LINUX Boot Screen ...
Страница 212: ...D 2 Boot Screen Doppler Boot Screen ...
Страница 213: ...P N 046 009019 00 11 0 ...