3-2 System Installation
3.1.2
Installation Condition
3.1.2.1
Space Requirements
Place the system with the necessary accessories at a proper position for convenient
use.
1. Place the system in a room with good ventilation or having an air conditioning
unit.
2. Leave at least 20cm clearance around the system to ensure effective cooling.
3. A combination lighting system in the room (dim/bright) is recommended.
4. Except the receptacle dedicated for the ultrasound system, at least 3-4 spare
receptacles on the wall are available for the other medical devices and peripheral
devices.
5. Power outlet and place for any external peripheral are within 2 m of each other
with peripheral within 1 m of the unit to connect cables.
3.1.2.2
Network Environment
Wired LAN function is supported by this ultrasound system.
Data transmission is allowed between different departments or areas with network
cable. Confirm the network devices and network conditions before the installation:
1. General information: default gateway IP address, and the other routers related
information.
2. DICOM application information: DICOM server name, DICOM port, channels, and
IP address.
3.1.2.3
Confirmation before Installation
Please confirm the following items before installation:
1. The video format used in the region or country where the system is installed.
2. The language used in the region or country where the system is installed.
3. The power voltage and frequency used in the region or country where the system
is installed.
4. Obstetric formulae and other measurement formulae used in the region or
country where the system is installed.
5. Other settings to be used in the region or country where the system is installed
but different from the factory settings.
6.
The doctor’s habits when using the system.
Please confirm the items above prior to the installation training, and do the system
settings according to the universal setting of installed region or country.
Содержание DC-25
Страница 1: ...DC 32 DC 30 DC 28 DC 26 DC 25 Digital Ultrasound System Service Manual Revision 11 0 ...
Страница 2: ......
Страница 7: ...v Appendix C Requirements of Performance Indices C 1 Appendix D Boot Screen D 1 ...
Страница 8: ......
Страница 12: ......
Страница 70: ......
Страница 88: ...5 18 Function Checking and Testing ...
Страница 94: ...6 6 Software installation and Maintenance ...
Страница 168: ......
Страница 178: ......
Страница 190: ......
Страница 206: ......
Страница 211: ...Boot Screen D 1 Appendix D Boot Screen BIOS Boot Screen LINUX Boot Screen ...
Страница 212: ...D 2 Boot Screen Doppler Boot Screen ...
Страница 213: ...P N 046 009019 00 11 0 ...