Options Installation
8-1
8
Options Installation
8.1
Installation of Optional Devices
1. Copy the optional devices to the U disk, then insert it to the USB port of the ultrasound
system;
2. Enter [Preset], click [Maintenance] to enter [Option] menu to select the option module to
be installed, click [Install].
3.
Select the corresponding key file in the prompted “Load File” dialogue box and click [OK].
4. After the Key file is successfully installed
, you can see the option module in the “Installed”
status. Corresponding functions are available after return from preset.
Note:
After all modules are installed, please go to the previous interface to confirm.
Trial: select the corresponding software package, and then click [Trial].
Note
: for each component, you can activate trial version only once. The trial lasts 180 days for each
key.
Promote
Note:
The service engineers must log in the system with the account of
Service
, and
then conduct the promote function.
Содержание DC-25
Страница 1: ...DC 32 DC 30 DC 28 DC 26 DC 25 Digital Ultrasound System Service Manual Revision 11 0 ...
Страница 2: ......
Страница 7: ...v Appendix C Requirements of Performance Indices C 1 Appendix D Boot Screen D 1 ...
Страница 8: ......
Страница 12: ......
Страница 70: ......
Страница 88: ...5 18 Function Checking and Testing ...
Страница 94: ...6 6 Software installation and Maintenance ...
Страница 168: ......
Страница 178: ......
Страница 190: ......
Страница 206: ......
Страница 211: ...Boot Screen D 1 Appendix D Boot Screen BIOS Boot Screen LINUX Boot Screen ...
Страница 212: ...D 2 Boot Screen Doppler Boot Screen ...
Страница 213: ...P N 046 009019 00 11 0 ...