5-7
5.6
Pausing Alarms
You can temporarily disable alarm indicators by pressing the
hardkey on the equipment’s front. When alarms are
paused:
For physiological alarms, no alarm indication is shown. New physiological alarm will not be presented.
The remaining alarm pause time is displayed in the physiological alarm area.
For technical alarms, alarm sounds are paused, but alarm lamps and alarm messages remain presented.
is displayed in the sound symbol area. If a new technical alarm is triggered in the alarm paused period, the
alarm message will be displayed.
When the alarm pause time expires, the alarm paused status is automatically deactivated. You can also cancel the alarm
paused status by pressing the
hardkey.
The default alarm pause time is 2 minutes. To change alarm pause time,
1.
Press the Main Menu button on the front panel, and then select [
Others >>
]
[
→
Configuration >>
]
enter the
→
required password.
2. Select
[
Alarm Setup >>
]
[
→
Alarm Pause Time
] and then select an appropriate value.
5.7
Switching Alarms Off
When an alarm is switched off, the alarm status is the same with that when an alarm is paused.
Alarms are switched off when:
The
hardkey is pressed if the [
Alarm Pause Time
] is set to [
Permanent
].
The equipment is switched to the Manual Defib mode. Or
The equipment exit synchronous defibrillation when operating in the Manual Defib mode.
The alarm off status is exited when,
The
hardkey is pressed, or
Sync Defib is switched on in Manual Defib mode.
Содержание BeneHeart D3
Страница 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES...
Страница 16: ...8 FOR YOUR NOTES...
Страница 22: ...1 6 FOR YOUR NOTES...
Страница 25: ...2 3 2 3 Main Unit 2 3 1 Front View Area 1 Alarm lamp Area 2 External paddle Handle Area 3 Microphone Speaker...
Страница 40: ...4 2 FOR YOUR NOTES...
Страница 62: ...6 12 FOR YOUR NOTES...
Страница 78: ...8 10 FOR YOUR NOTES...
Страница 88: ...10 4 FOR YOUR NOTES...
Страница 90: ...11 2 FOR YOUR NOTES...
Страница 102: ...14 2 FOR YOUR NOTES...
Страница 170: ...D 6 FOR YOUR NOTES...
Страница 182: ...G 2 FOR YOUR NOTES...
Страница 184: ......
Страница 185: ......
Страница 186: ...P N 046 001653 00 6 0...