22-4
22.3.4
User Test
WARNING
z
Do not perform user test when a patient is connected to the equipment.
User test covers the following items:
Routine Test,
Energy delivery test, and
Controls test.
NOTE
z
Before user test or after each use, thoroughly clean the paddles and properly place them in the paddle tray.
User test passes only when paddles properly contact the metal parts of the paddle tray.
z
Install the battery and properly place the external paddles in the paddle tray or connect the pads cable and
50 Ω test load. Otherwise the User Test will fail.
22.3.4.1
Entering the User Test Main Menu
To access user test, press the Main Menu button and select [
User Test >>
]. Then a dialog box pops up, prompting you
that entering user test will end patient monitoring. Select [
Yes
] to enter the User Test Main menu.
Check the test items you want to perform and select [
Start
] to start user test. The message “Test completed” will be
presented when selected tests have been finished. Then you can press the [
Record
] soft key to print the test result.
Содержание BeneHeart D3
Страница 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES...
Страница 16: ...8 FOR YOUR NOTES...
Страница 22: ...1 6 FOR YOUR NOTES...
Страница 25: ...2 3 2 3 Main Unit 2 3 1 Front View Area 1 Alarm lamp Area 2 External paddle Handle Area 3 Microphone Speaker...
Страница 40: ...4 2 FOR YOUR NOTES...
Страница 62: ...6 12 FOR YOUR NOTES...
Страница 78: ...8 10 FOR YOUR NOTES...
Страница 88: ...10 4 FOR YOUR NOTES...
Страница 90: ...11 2 FOR YOUR NOTES...
Страница 102: ...14 2 FOR YOUR NOTES...
Страница 170: ...D 6 FOR YOUR NOTES...
Страница 182: ...G 2 FOR YOUR NOTES...
Страница 184: ......
Страница 185: ......
Страница 186: ...P N 046 001653 00 6 0...