Open Channel Monitoring (OCM)
OCM converts a level reading (head) into a flow value using a linear algorithm. When the
MiniRanger is used to calculate flow the mA Output is scaled to the maximum flow (P604)
rather than the span (P007).
Example Flow Characterization
Program the parameters below to define the maximum and minimum points on the curve.
Then program the two following parameters to closely match the flow curve of your
primary measuring device (PMD).