Electrical connection
Local installation
;
The
sintering oven
requires its own electrical circuit.
;
The building's electrical circuit must have a type-K or type-Z circuit
breaker with a rated current of at least 16 A (other types of circuit
breaker depending on the country of use).
;
An additional residual-current circuit breaker (designed for 30 mA
tripping current) must be installed.
;
To ensure electrically safe operation, the
sintering oven
requires a
protective earth conductor to be connected to the power socket.
;
When selecting the installation location, ensure that the accompany-
ing mains cable is 2.0 m long. Extending the cable is not permitted.
The supply voltage must be within the rated voltage range of 200-240
V.
Danger
Electricity!
Risk of death from electric shock.
h
Do not touch live cables and components with wet hands.
h
Observe the accident-prevention regulations when working with elec-
tric current.
h
Only connect the unit to a voltage supply that matches the specifica
-
tions on the rating plate.
Warning
Release of pollutants!
h
Suitable respiratory protection must be worn when handling insulating
material.
h
If necessary, an extraction system must be installed.