Appendix | EtherCAT Documentation
capaNCDT 6500
A 5.3
Measurement Data Format
The measuring values are transmitted as Signed32.
Since the controller has a resolution of 24 bit, not all 32 bits are required. 0x0 = 0 % of
measuring range corresponds consequently to 0xFFFFFF = 100 % of measuring range.
The measuring range can be read from the channel info objects 2020h to 2027h (range
and unit). Here is also the minimum and maximum value, which the Signed32 measuring
value can take (data format zero value and data format end value).
A 5.4
EtherCAT Configuration with the Beckhoff TwinCAT©-Manager
For example the Beckhoff TwinCAT Manager can be used as EtherCAT Master.
Copy the device description file (EtherCAT®-Slave Information)
Micro-Epsilon.
xml
from the included CD in the directory \\TwinCAT\IO\EtherCAT before the measu-
ring device can be configured via EtherCAT®.
EtherCAT®-Slave information files are XML files, which specify the characteristics of the
Slave device for the EtherCAT® Master and contain information to the supported com-
munication objects.
Restart the TwinCAT Manager after copying.
Searching for a device:
Select the tab
I/O Devices,
then
Scan Devices
.
Confirm with
OK.
Select a network card, where EtherCAT®–Slaves should be searching for.
Confirm with
OK.
The window
Search for new boxes
(EtherCAT®-Slaves) appears.
Confirm with
Yes.
The DT6530 is now shown in a list.
Now confirm the window
Activate Free Run
with
Yes.