Ethernet Interface
capaNCDT 6500
Page 39
6.4.9 Mode of Linearization (LIN)
Specifies the linearization type for each channel.
The linearization type can be set for each channel. The index m stands for channel num-
ber, the index n for the linearization type.
LIN
Command
$LINm:n<CR> (for example: $LIN5:2<CR> = 2-point-linearization for
channel 5)
Response
$LINm:nOK<CRLF>
Index m
(Channel num-
ber)
1 … 8
Index n (lineari-
zation mode)
0 = no linearization (Default setting)
1 = Start of measuring range
2 = 2-point-linearization
3 = 3-point-linearization
4 = 5-point-linearization
5 = 10-point-linearization
Request linearization mode
Command
$LIN?<CR>
Response
$LIN?n,n,n,n,n,n,n,nOK<CRLF> (n stands for the linearization type)
6.4.10 Set Linearization Point (SLP)
Sets a linearization point.
Place the sensor or the target in the corresponding position. After the command has
been received, the current measuring value will be recorded at this position as a lineari-
zation point and thus the constants are recalculated for the linearization.
SLP
Command
$SLPm:n<CR> (for example: $SLP5:3<CR> =linearization point at
30 % of channel 5)
Response
$SLPm:nOK<CRLF>
Index m
(Channel num-
ber)
1 ... 8
Index n (point of
linearization)
n (linearization point):
1 = linearization point at 10 % of the measuring range
2 = linearization point at 20 % of the measuring range
3 = linearization point at 30 % of the measuring range
4 = linearization point at 40 % of the measuring range
5 = linearization point at 50 % of the measuring range
6 = linearization point at 60 % of the measuring range
7 = linearization point at 70 % of the measuring range
8 = linearization point at 80 % of the measuring range
9 = linearization point at 90 % of the measuring range
10 = linearization point at 100 % of the measuring range