Ethernet Interface
capaNCDT 6500
Page 33
Select
Internet Protocol (TCP/IP)
>
Properties.
i
In order to be able to use the Ethernet interface, there must be a demodulator mod-
ule in channel 1 as this one determines the cycle for all channels!
By default, the IP address of the controller is set to 169.254.168.150. Communication
with the controller is done on the data port 10001 for measurement transmission. A com-
mand port (Telnet, port 23) is used for sensor commands.
The IP settings and the data port can be changed at any time:
-
by using the web browser. Enter the current IP address into the address bar. Go to the
menu
Settings
>
Digital Interfaces
>
Ethernet settings
to set a new IP
address, activate DHCP or change the data port.
-
by using software commands, see Chap.
6.4
.
-
by using the
SensorFinder
software.
If you activate DHCP, access to the controller via a DHCP host name is possible. The
host name contains the device name and serial number. Structure: NAME_SN e. g.
DT6530_1001.
The controller supports UPnP. If you use an operational system with activated UPnP
client e. g. standard with Windows 7, the controller is listed in the explorer as a device
automatically. This is helpful, if you do not know the IP address of the controller.