QMA401 User’s Manual
42
97450 Issue 2, March 2018
OPERATION
3.9
Measurement Cycle
At the beginning of a measurement cycle V1 is energized. This allows the dried sample
gas to be routed to the sensor cell for a period of 30 seconds as shown by the red line
in
Figure 25. During this first phase of the measurement cycle the frequency difference
between the sensor and reference crystals is measured (i.e. dry state).
The sample and calibration gas paths are shown in green. These lines are continually
purged during the initial measurement cycle.
SAMPLE
IN
SAMPLE
OUT
DD
HE1
HE2
FC1
V1
V2
V3
MFC
QCH
PS
MG
FC2
FC3
Key
DD
Desiccant column
MG
Moisture generator
MFC
Mass flow controller
V1, V2, V3
Solenoid valves
QCH
Sensor cell
HE1, HE2
Heat exchanger
PS
Pressure sensor
FC1, FC2, FC3
Flow control
Figure 38
Measurement Cycle (Phase 1) - Dried Sample Flow
Содержание QMA401
Страница 1: ...QMA401 Trace Moisture Analyzer User s Manual 97450 Issue 2 March 2018...
Страница 60: ...QMA401 User s Manual 50 97450 Issue 2 March 2018 CALIBRATION Figure 42 Typical QMA401 Calibration Certificate...
Страница 71: ...QMA401 User s Manual Michell Instruments 61 APPENDIX A Appendix A Technical Specifications...
Страница 73: ...QMA401 User s Manual Michell Instruments 63 APPENDIX B Appendix B Calculating Conversion Factors for Gas Mixes...
Страница 75: ...QMA401 User s Manual Michell Instruments 65 APPENDIX C Appendix C Modbus Register Map...
Страница 95: ...QMA401 User s Manual Michell Instruments 85 APPENDIX D Appendix D Quality Recycling Warranty Information...
Страница 97: ...QMA401 User s Manual Michell Instruments 87 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 99: ......
Страница 100: ...QMA401 User s Manual 90 97450 Issue 2 March 2018 NOTES...
Страница 101: ...QMA401 User s Manual Michell Instruments 91 NOTES...
Страница 102: ...http www michell com...