![Michell Instruments QMA401 Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/michell-instruments/qma401/qma401_user-manual_1784687078.webp)
QMA401 User’s Manual
68
97450 Issue 2, March 2018
APPENDIX C
81
*Osc Table1 DeltaF 10 Lo Word
R/W
I
“
82
*Osc Table1 DeltaF 11 Hi Word
R/W
I
0.0001 to 2000.0000
83
*Osc Table1 DeltaF 11 Lo Word
R/W
I
“
84
*Osc Table1 DeltaF 12 Hi Word
R/W
I
0.0001 to 2000.0000
85
*Osc Table1 DeltaF 12 Lo Word
R/W
I
“
86
*Osc Table1 Ref 01 Hi Word
R/W
I
0.0001 to 10000.0000
87
*Osc Table1 Ref 01 Lo Word
R/W
I
“
88
*Osc Table1 Ref 02 Hi Word
R/W
I
0.0001 to 10000.0000
89
*Osc Table1 Ref 02 Lo Word
R/W
I
“
90
*Osc Table1 Ref 03 Hi Word
R/W
I
0.0001 to 10000.0000
91
*Osc Table1 Ref 03 Lo Word
R/W
I
“
92
*Osc Table1 Ref 04 Hi Word
R/W
I
0.0001 to 10000.0000
93
*Osc Table1 Ref 04 Lo Word
R/W
I
“
94
*Osc Table1 Ref 05 Hi Word
R/W
I
0.0001 to 10000.0000
95
*Osc Table1 Ref 05 Lo Word
R/W
I
“
96
*Osc Table1 Ref 06 Hi Word
R/W
I
0.0001 to 10000.0000
97
*Osc Table1 Ref 06 Lo Word
R/W
I
“
98
*Osc Table1 Ref 07 Hi Word
R/W
I
0.0001 to 10000.0000
99
*Osc Table1 Ref 07 Lo Word
R/W
I
“
100
*Osc Table1 Ref 08 Hi Word
R/W
I
0.0001 to 10000.0000
101
*Osc Table1 Ref 08 Lo Word
R/W
I
“
102
*Osc Table1 Ref 09 Hi Word
R/W
I
0.0001 to 10000.0000
103
*Osc Table1 Ref 09 Lo Word
R/W
I
“
104
*Osc Table1 Ref 10 Hi Word
R/W
I
0.0001 to 10000.0000
105
*Osc Table1 Ref 10 Lo Word
R/W
I
“
106
*Osc Table1 Ref 11 Hi Word
R/W
I
0.0001 to 10000.0000
107
*Osc Table1 Ref 11 Lo Word
R/W
I
“
108
*Osc Table1 Ref 12 Hi Word
R/W
I
0.0001 to 10000.0000
109
*Osc Table1 Ref 12 Lo Word
R/W
I
“
110
*BLANK3
R/W
Don’t Care
111
*Osc Table1 Cal Flow Rate ml/m
R/W
A2
10-2000 ml/m
112
*Osc Table1 Cal Int Moist Gen Val Hi Word
R/W
I
0.0001 to 10000.0000
113
*Osc Table1 Cal Int Moist Gen Val Lo Word
R/W
I
“
114
*Osc Table1 Cal Oven SP degC/ Cal Date
DAY
R/W
Q
40 to 80 / 1 to 31
115
*Osc Table1 Cal Date MONTH / YEAR
R/W
Q
1 to 12 / 0 to 99
116
*Osc Table1 Cal Sample Phase Time
R/W
A1
10 to 65535 seconds
117
*Osc Table1 Cal Reference Phase Time
R/W
A1
10 to 65535 seconds
118
*Osc Table1 Cal Settling Cycles
R/W
A1
4 to 240 cycles
119
*Osc Table1 Cal Calibration Cycles
R/W
A1
4 to 60 cycles
120
*BLANK
R/W
Don’t Care
121
*Osc Table1 Cal Cell Pressure Reading
R/W
A3
0.00 to 10.00 barG
122
*Osc Table1 Cal Beat Freq of Ref
R/W
A1
1000 to 15000 (NOT
USED)
123
*Osc Table1 Cal Beat Freq of Moist Gen
R/W
A1
1000 to 15000 (NOT
USED)
124
*Osc Table2 DeltaF 01 Hi Word
R/W
I
0.0001 to 2000.0000
125
*Osc Table2 DeltaF 01 Lo Word
R/W
I
“
126
*Osc Table2 DeltaF 02 Hi Word
R/W
I
0.0001 to 2000.0000
127
*Osc Table2 DeltaF 02 Lo Word
R/W
I
“
128
*Osc Table2 DeltaF 03 Hi Word
R/W
I
0.0001 to 2000.0000
129
*Osc Table2 DeltaF 03 Lo Word
R/W
I
“
Содержание QMA401
Страница 1: ...QMA401 Trace Moisture Analyzer User s Manual 97450 Issue 2 March 2018...
Страница 60: ...QMA401 User s Manual 50 97450 Issue 2 March 2018 CALIBRATION Figure 42 Typical QMA401 Calibration Certificate...
Страница 71: ...QMA401 User s Manual Michell Instruments 61 APPENDIX A Appendix A Technical Specifications...
Страница 73: ...QMA401 User s Manual Michell Instruments 63 APPENDIX B Appendix B Calculating Conversion Factors for Gas Mixes...
Страница 75: ...QMA401 User s Manual Michell Instruments 65 APPENDIX C Appendix C Modbus Register Map...
Страница 95: ...QMA401 User s Manual Michell Instruments 85 APPENDIX D Appendix D Quality Recycling Warranty Information...
Страница 97: ...QMA401 User s Manual Michell Instruments 87 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 99: ......
Страница 100: ...QMA401 User s Manual 90 97450 Issue 2 March 2018 NOTES...
Страница 101: ...QMA401 User s Manual Michell Instruments 91 NOTES...
Страница 102: ...http www michell com...