QMA401 User’s Manual
Michell Instruments
79
APPENDIX C
Register Configuration M — System Warning Flags (Register 230), Analyzer Status
Alarm Relay Selection Mask (Register 9)
1=Warning or fault, 0=OK
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
Bit
HEX
Warning Condition
0
0001
Oven temperature unstable.
Oven temperature has not been
stable within ±0.01C of setpoint for continuous 5 minutes. (Process
alarms de-energized, both analog outputs at fault condition
1
0002
Enclosure temperature too high.
Enclosure (System) temperature
too high. (> Oven temperature setpoint -2°C)
2
0004
Flow control error.
MFC Flow control error (by >5ml/m of fl ow
target)
3
0008
Cell pressure sensor error.
(under 4 mA, over 20 mA or no signal)
4
0010
Ext Press sensor error.
(under 4 mA, over 20 mA or no signal)
5
0020
Field cal error.
Internal moist generator drift, instrument excessive
drift or desiccant dryer deterioration requiring very large corr. factor
(<0.2500 or >4.000). In this case the correction factor would be set
to 1.0. (Checked after fi eld calibration)
6
0040
Beat freq. out of range.
Beat frequency Under/Over acceptable
range (<1500 Hz, >20000 Hz)
7
0080
ppmV over range.
ppm
V
over instrument range (>2000ppm
V
)
8
0100
Oven temperature sensor fault.
Oven temperature sensor fault
(Process Alarms de-energized, Fault Alarm Active, both Analog
outputs at fault condition (ADC <10, >4000 counts)
9
0200
mA output 1 error.
(o/c or high resistance at output)
10
0400
mA output 2 error.
(o/c or high resistance at output)
11
0800
Oscillator board comms error.
Oscillator board comms. error or
board not present (checked on startup)
12
1000
Ethernet board comms. error.
Ethernet Board comms. error or
board not fi tted (checked on startup)
13
2000
Dryer due for service.
Desiccant Dryer due for service or
replacement (>5000000ppm
V
)
14
4000
Moisture generator due for service.
Internal Moisture Generator
due for service or replacement (>1030days)
15
8000
Calibration Alarm
In calibration mode or data held (or both)
Analyzer Status Alarm selection mask (in Register 9) allows user to set which condition(s) trigger
status relay alarm 4.
Содержание QMA401
Страница 1: ...QMA401 Trace Moisture Analyzer User s Manual 97450 Issue 2 March 2018...
Страница 60: ...QMA401 User s Manual 50 97450 Issue 2 March 2018 CALIBRATION Figure 42 Typical QMA401 Calibration Certificate...
Страница 71: ...QMA401 User s Manual Michell Instruments 61 APPENDIX A Appendix A Technical Specifications...
Страница 73: ...QMA401 User s Manual Michell Instruments 63 APPENDIX B Appendix B Calculating Conversion Factors for Gas Mixes...
Страница 75: ...QMA401 User s Manual Michell Instruments 65 APPENDIX C Appendix C Modbus Register Map...
Страница 95: ...QMA401 User s Manual Michell Instruments 85 APPENDIX D Appendix D Quality Recycling Warranty Information...
Страница 97: ...QMA401 User s Manual Michell Instruments 87 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 99: ......
Страница 100: ...QMA401 User s Manual 90 97450 Issue 2 March 2018 NOTES...
Страница 101: ...QMA401 User s Manual Michell Instruments 91 NOTES...
Страница 102: ...http www michell com...