QMA401 User’s Manual
62
97450 Issue 2, March 2018
APPENDIX A
Appendix A
Technical Specification
Performance
Measurement
Technology
Fast Response Quartz Crystal Microbalance
Calibrated Range
0.1 to 700ppm
V
Measurement Range
0.1 to 2000ppm
V
Accuracy
±10% of reading from 1 to 2000 ppm
V
±0.1 ppm
V
between 0.1 & 1 ppm
V
Repeatability
±5% of the reading from 1 to 2000 ppm
V
±0.1 ppm
V
between 0.1 & 1 ppm
V
Detection Limit
0.1 ppm
V
Available Units
ppm
V
, ppm
W
, mg/m
3
, vapor pressure (Pa), frost point (°C), lbs/MMscf
Response Speed
T63 <2 mins to step change in either direction
T95 <5 mins to step change in either direction
Automatic Calibration
Internal moisture generator source calibrated traceable to NPL & NIST
Sensitivity
0.1
ppm
V
or 1% of reading, whichever is greater
Electrical Specifications
Supply Voltage
85 to 264 V AC, 47/63Hz, 110 to 300 V DC
Max Power
Consumption
111W
Alarms
1 x System Alarm, volt-free change-over (FORM C)
3 x process alarms, selectable for various parameters, volt free change-
over (FORM C)
Communications
Analog Outputs: 2 channels, user selectable 4-20 mA or 1 to 5 V
Digital Communications: RS485/USB Modbus RTU, Ethernet Modbus TCP
Data logging
Logging to SD card at user-selectable interval or to PC via application software
Local Interface
7” color touch screen LCD
Operating Conditions
Inlet Pressure
1 barg max (14.5psig)
With optional inlet pressure regulator: 300 barg max (4351psig)
Outlet Pressure
Atmospheric
Sample Flow
300ml/min total flow without bypass
Sample Gas
Temperature
0 to +100°C
Operating Envir5 to +45°C (+41 to +113°F) up to 90% RH
Mechanical Specifications
Enclosure
19" Rackmount, 4U x 434 mm
Gas Connections
1/4” VCR(M)
Weight
13.5 kg (29.8 lbs)
Содержание QMA401
Страница 1: ...QMA401 Trace Moisture Analyzer User s Manual 97450 Issue 2 March 2018...
Страница 60: ...QMA401 User s Manual 50 97450 Issue 2 March 2018 CALIBRATION Figure 42 Typical QMA401 Calibration Certificate...
Страница 71: ...QMA401 User s Manual Michell Instruments 61 APPENDIX A Appendix A Technical Specifications...
Страница 73: ...QMA401 User s Manual Michell Instruments 63 APPENDIX B Appendix B Calculating Conversion Factors for Gas Mixes...
Страница 75: ...QMA401 User s Manual Michell Instruments 65 APPENDIX C Appendix C Modbus Register Map...
Страница 95: ...QMA401 User s Manual Michell Instruments 85 APPENDIX D Appendix D Quality Recycling Warranty Information...
Страница 97: ...QMA401 User s Manual Michell Instruments 87 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 99: ......
Страница 100: ...QMA401 User s Manual 90 97450 Issue 2 March 2018 NOTES...
Страница 101: ...QMA401 User s Manual Michell Instruments 91 NOTES...
Страница 102: ...http www michell com...