QMA401 User’s Manual
Michell Instruments
13
OPERATION
3.2.1 Regulating Sensor Pressure
The sensor pressure of the QMA401 must match the pressure stated on the calibration
certificate in order for the calibration to be valid.
To control the sensor pressure, a pressure regulator is required on the input of the
analyzer.
An integral pressure regulator (rated to 300barg input pressure) can be supplied as an
option at time of order. Otherwise, when choosing an external pressure regulator, select
one designed for use in high purity gas, with a stainless steel body and diaphragm to
minimize impact on sample response speed when measuring dry samples.
To set the input pressure:
1.
Power the analyzer on and wait for the oven to heat up before proceeding.
2.
Navigate to the Monitor Screen.
Sensor pressure
shows the actual
pressure seen by the cell.
3.
Use the pressure regulator to adjust the
Sensor pressure
to match the
pressure shown on your calibration certificate.
NOTE: The pressure at the outlet should be atmospheric, unless
otherwise stated on the calibration certificate.
4.
Familiarize yourself with the menu system of the analyzer, and setup the
parameters listed in the next section. Before a measurement is made,
ensure the correct carrier gas has been selected in the menu system.
Содержание QMA401
Страница 1: ...QMA401 Trace Moisture Analyzer User s Manual 97450 Issue 2 March 2018...
Страница 60: ...QMA401 User s Manual 50 97450 Issue 2 March 2018 CALIBRATION Figure 42 Typical QMA401 Calibration Certificate...
Страница 71: ...QMA401 User s Manual Michell Instruments 61 APPENDIX A Appendix A Technical Specifications...
Страница 73: ...QMA401 User s Manual Michell Instruments 63 APPENDIX B Appendix B Calculating Conversion Factors for Gas Mixes...
Страница 75: ...QMA401 User s Manual Michell Instruments 65 APPENDIX C Appendix C Modbus Register Map...
Страница 95: ...QMA401 User s Manual Michell Instruments 85 APPENDIX D Appendix D Quality Recycling Warranty Information...
Страница 97: ...QMA401 User s Manual Michell Instruments 87 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 99: ......
Страница 100: ...QMA401 User s Manual 90 97450 Issue 2 March 2018 NOTES...
Страница 101: ...QMA401 User s Manual Michell Instruments 91 NOTES...
Страница 102: ...http www michell com...