QMA401 User’s Manual
Michell Instruments
21
OPERATION
3.6.3 Alarm Screen
Figure 16
Alarm Screen
Parameter
Description
Alarm Selector
The left and right arrow keys are used to toggle through the different
alarms available.
Available Options:
Alarm 1, Alarm 2, Alarm 3, Alarm 4 - Analyzer
Status Alarm only
Clear Latch
Clears a latched alarm.
Available Options:
Yes, Cleared
Parameter
Selects the parameter for the corresponding alarm.
Available Options:
Moisture Content ppm
V
Moisture Content ppm
W
Moisture Content mg/m
3
Water Vapor Pressure Pa
Moisture Content lbs/MMscf
Frost Point
Oven Temperature
Flow ml/min
Cell Pressure
External Pressure
Low
Selects the lower alarm limit for the corresponding alarm using the
keypad which is opened.
Latch
Selects between latched and non-latched alarms.
Available Options:
On, Off
High
Selects the higher alarm limit for the corresponding alarm using the
keypad which is opened.
Table 2
Alarm Screen Parameters
Содержание QMA401
Страница 1: ...QMA401 Trace Moisture Analyzer User s Manual 97450 Issue 2 March 2018...
Страница 60: ...QMA401 User s Manual 50 97450 Issue 2 March 2018 CALIBRATION Figure 42 Typical QMA401 Calibration Certificate...
Страница 71: ...QMA401 User s Manual Michell Instruments 61 APPENDIX A Appendix A Technical Specifications...
Страница 73: ...QMA401 User s Manual Michell Instruments 63 APPENDIX B Appendix B Calculating Conversion Factors for Gas Mixes...
Страница 75: ...QMA401 User s Manual Michell Instruments 65 APPENDIX C Appendix C Modbus Register Map...
Страница 95: ...QMA401 User s Manual Michell Instruments 85 APPENDIX D Appendix D Quality Recycling Warranty Information...
Страница 97: ...QMA401 User s Manual Michell Instruments 87 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 99: ......
Страница 100: ...QMA401 User s Manual 90 97450 Issue 2 March 2018 NOTES...
Страница 101: ...QMA401 User s Manual Michell Instruments 91 NOTES...
Страница 102: ...http www michell com...