OPERATION
3.8
Sampling Guidelines
The QMA401 Trace Moisture Analyzer is designed to operate in a flowing gas stream
and is suitable for the measurement of the moisture content of a wide variety of gases.
In general, if the gas (in conjunction with water vapor) is not corrosive to the sampling
system and the sensor base metals then it will be suitable for measurement by the
QMA401.
The analyzer is designed to automatically regulate the flow rate. However, the sample
pressure and back pressure must match what is shown on the calibration certificate -
typically 1 barg (14.5 psig) sample pressure with the oulet at atmospheric pressure, and
should be controlled using a high-quality pressure regulator on the gas inlet and a back
pressure regulator on the outlet.
General guidelines to be followed when setting-up a sampling system are as follows:
•
Ensure that the sample is representative of the gas under test
To ensure that the sample is representative of the process being monitored, the sample
point should be as close to the critical measurement point as possible. Also, never
sample from the bottom of a pipe where entrained liquids may be drawn into the
instrument’s sample input line.
•
Minimize the ‘dead space’ in sample lines
Dead space in sample lines causes moisture entrapment points, increased system
response times or measurement errors as the trapped moisture is released into passing
sample gas, producing an increase in partial vapor pressure.
Avoid the use of too many T-pieces, in-line couplings or other unnecessary pipework.
Sample pipework should, ideally, be specially designed for each application rather than
adapted from that previously installed for another application. Dead space in sample
lines increases response time by holding water molecules which are more slowly released
to the passing gas sample.
•
Remove any particulate matter or oil from the gas sample
Particulate matter can damage the sensors. If particulate matter, such as degraded
desiccant, pipe scale and rust are likely to be present in the sample gas, use a particulate
in-line filter. Michell Instruments’ technical sales department can be contacted for advice.
Содержание QMA401
Страница 1: ...QMA401 Trace Moisture Analyzer User s Manual 97450 Issue 2 March 2018...
Страница 60: ...QMA401 User s Manual 50 97450 Issue 2 March 2018 CALIBRATION Figure 42 Typical QMA401 Calibration Certificate...
Страница 71: ...QMA401 User s Manual Michell Instruments 61 APPENDIX A Appendix A Technical Specifications...
Страница 73: ...QMA401 User s Manual Michell Instruments 63 APPENDIX B Appendix B Calculating Conversion Factors for Gas Mixes...
Страница 75: ...QMA401 User s Manual Michell Instruments 65 APPENDIX C Appendix C Modbus Register Map...
Страница 95: ...QMA401 User s Manual Michell Instruments 85 APPENDIX D Appendix D Quality Recycling Warranty Information...
Страница 97: ...QMA401 User s Manual Michell Instruments 87 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 99: ......
Страница 100: ...QMA401 User s Manual 90 97450 Issue 2 March 2018 NOTES...
Страница 101: ...QMA401 User s Manual Michell Instruments 91 NOTES...
Страница 102: ...http www michell com...