3.10 Calibration Cycle
To maintain the precision of the analyzer, the unit can self calibrate and adjust its
internal reference table based on the result.
This is achieved as follows:
An internal moisture generator uses a permeation tube to generate a nominal moisture
content of 0.5, 5 or 50 ppm
V
, depending on what was specified at the time of order.
The calibration is carried out in a two phase cycle. Initially V1 is energized, causing the
dried sample gas to be routed to the sensor cell for a period of 30 seconds, as shown
by the red line in
Figure 27.
SAMPLE
IN
SAMPLE
OUT
DD
HE1
HE2
FC1
V1
V2
V3
MFC
QCH
PS
MG
FC2
FC3
Key
DD
Desiccant column
MG
Moisture generator
MFC
Mass flow controller
V1, V2, V3
Solenoid valves
QCH
Sensor cell
HE1, HE2
Heat exchanger
PS
Pressure sensor
FC1, FC2, FC3
Flow control
Figure 40
Calibration Cycle (Phase 1) - Dried Sample Flow
Содержание QMA401
Страница 1: ...QMA401 Trace Moisture Analyzer User s Manual 97450 Issue 2 March 2018...
Страница 60: ...QMA401 User s Manual 50 97450 Issue 2 March 2018 CALIBRATION Figure 42 Typical QMA401 Calibration Certificate...
Страница 71: ...QMA401 User s Manual Michell Instruments 61 APPENDIX A Appendix A Technical Specifications...
Страница 73: ...QMA401 User s Manual Michell Instruments 63 APPENDIX B Appendix B Calculating Conversion Factors for Gas Mixes...
Страница 75: ...QMA401 User s Manual Michell Instruments 65 APPENDIX C Appendix C Modbus Register Map...
Страница 95: ...QMA401 User s Manual Michell Instruments 85 APPENDIX D Appendix D Quality Recycling Warranty Information...
Страница 97: ...QMA401 User s Manual Michell Instruments 87 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 99: ......
Страница 100: ...QMA401 User s Manual 90 97450 Issue 2 March 2018 NOTES...
Страница 101: ...QMA401 User s Manual Michell Instruments 91 NOTES...
Страница 102: ...http www michell com...