3.2 Remote control commands
719 S Titrino
89
I
nfo
.A
ctualInfo
.D
isplay
.L2
up to 24 ASCII characters
Lines of the display. The display can be written to from the computer. Pro-
ceed as follows:
Lock the display, see 3.2.2.80.
$Q sends the contents of the corresponding display line.
3.2.2.71. I
nfo
.As
sembly
.C
ycleTime read
only
I
nfo
.As
sembly
.E
xV read
only
Inquiries regarding basic variables of the assembly: Cycle time in s, volume of
the active Exchange Unit in mL.
3.2.2.72. A
ssembly
.B
ur
.R
ates
.F
orward
.S
elected digital,
analog
A
ssembly
.B
ur
.R
ates
.F
orward
.D
igital 0...150,
max.
A
ssembly
.B
ur
.R
ates
.R
everse
.S
elected digital,
analog
A
ssembly
.B
ur
.R
ates
.R
everse
.D
igital 0...150,
max.
Expel and aspirating rate.
Digital or analog control. With digital control, the inputted value applies (in
mL/min). "max." means maximum possible rate with the Exchange Unit in
current use.
Analog means rate control with the analog potentiometer on Titrino.
3.2.2.73. A
ssembly
.B
ur
.F
ill
$G, $H, $C
$G starts the 'FILL' mode of the buret function.
3.2.2.74. A
ssembly
.B
ur
.M
odeDis
$G, $S, $H, $C
A
ssembly
.B
ur
.M
odeDis
.S
elected
volume
, time
A
ssembly
.B
ur
.M
odeDis
.V
0.0001...
0.1
...9999
A
ssembly
.B
ur
.M
odeDis
.T
ime 0.25...
1
...86400
A
ssembly
.B
ur
.M
odeDis
.VS
top 0.0001...9999,
OFF
A
ssembly
.B
ur
.M
odeDis
.A
utoFill ON,
OFF
Dispensing mode with parameters. The dispensing mode can only be started
and stopped via the RS Control. During a running dosification, no method can
be started at the Titrino.
.Selected: Dispensing of volume increments or during a preset time.
.Volume, .Time: Size of the volume increments or entry of time.
.VStop:
Limit volume for the dispensing.
.AutoFill: ON means automatic filling after every dispensing.
3.2.2.75. A
ssembly
.M
eas
.S
tatus ON,
OFF
A
ssembly
.M
eas
.M
easInput
1
, 2, Diff., Ipol, Upol, Temp
A
ssembly
.M
eas
.I
pol ±127...
1
...+127
A
ssembly
.M
eas
.U
pol ±1270...
400
...+1270
Measurement in assembly. The measuring function can only be started via RS
Control. When the measuring function is switched on, no method can be
started at the Titrino.
.Input:
Selection of the potentiometric measuring input 1, 2, diff., polar-
ized electrodes or temperature.
Содержание Titrino 719 S
Страница 2: ......
Страница 6: ...1 2 Controls and parts 719 S Titrino 2 1 2 Controls and parts Front view of instrument...
Страница 8: ...1 2 Controls and parts 719 S Titrino 4 Rear view of instrument...
Страница 58: ......
Страница 106: ......
Страница 135: ...6 2 Pin assignment of the Remote socket 719 S Titrino 131 6 2 1 Lines of the Remote socket during the titration...