4.2.
Diagnosis
719 S Titrino
115
Spindle remains at maximum position. The transit time of the spindle is
20 s.
•
Measure spindle lifting (can be performed only if the 3.496.0070 Dummy Exchange
Unit is inserted or the locking switch (in right hole) is carefully operated with a
screwdriver after removal of the Exchange Unit).
From the start point, the spindle travels 80 mm. Instead of the spindle height, the expelled
volume can be measured (corresponds to max. vol. of Exchange Unit used).
•
Press
<FILL>
(on instrument) and simultaneously measure the time until the Titrino
is again in the 'ready' position.
Times for filling:
per stop cock cycle 1 s
for filling
20 s (Tolerance: 10 %)
The following generally holds: Spindle and stopcock must move at a constant speed
(noise!). In the filling setting, the stopcock coupling must position the lever of the Exchange
Unit correctly at the left stop (with virtually no play and without sticking).
•
Set potentiometer '
dV/dt
' to left stop.
•
Press
<DOS>
(on instrument) at same time and use a stopwatch to measure the
time for 1
/
10 of the cylinder volume to be expelled. The time should be ca. 90 ... 110
s.
•
Set potentiometer '
dV/dt
' to right stop.
•
Press
<FILL>.
11 Setting up original arrangement
Reconnect all peripherals disconnected at the start of the diagnostic routine and perform
a short function test with these.
Содержание Titrino 719 S
Страница 2: ......
Страница 6: ...1 2 Controls and parts 719 S Titrino 2 1 2 Controls and parts Front view of instrument...
Страница 8: ...1 2 Controls and parts 719 S Titrino 4 Rear view of instrument...
Страница 58: ......
Страница 106: ......
Страница 135: ...6 2 Pin assignment of the Remote socket 719 S Titrino 131 6 2 1 Lines of the Remote socket during the titration...