4.2.
Diagnosis
719 S Titrino
111
Input 2
XX mV
•
Carry out the same measurements as with Input 1.
•
Press
<Clear>
.
analog input test 1...5
•
Short-circuit input ‘Ind I’ (e.g. with cable 3.496.5070).
•
Press
<3>
.
Input 1-2
XX mV
The differencial voltage between inputs ‘Ind I‘ and ‘Ind II’ is displayed.
Example: 0 - (+)1500 mV = -1500 mV
•
Remove cables from the inputs ‘Ind I’ and ‘Ind II’.
•
Press
<Clear>
.
analog input test 1...5
7.2 To check Pt 100 / 1000 connection
•
Connect a Pt 100 or Pt 1000 sensor, a resistor switch-box or a single resistor of 100
Ω
or 1 k
Ω
, respectively, to sockets ‘Pt 100/1000’ by means of short cables.
•
Press
<4>
.
Pt 100*
XX °C
(* or Pt 1000)
Tolerance: ± 0.5 °C (Note also tolerance of resistor switch-box).
The sensor allows automatic displaying of the room temperature. (The resistor values corre-
spond to 0°C.)
•
Press
<Clear>
.
analog input test 1...5
•
Remove cables and resistor switch-box.
7.3 Polarizer
test
•
Press
<5>
.
polarizer test
•
Press
<Enter>
.