2.7 Parameters, key <PARAM>
719 S Titrino
29
>preselections
conditioning: OFF
display drift: ON
cond.
drift corr: OFF
cond.
drift value 0.0
µ
l/min
cond.
req.ident: OFF
cond.
req.smpl size: OFF
cond.
activate pulse: OFF
cond.
Preselections for the sequence
Automatic conditioning of titration vessel. (ON, OFF)
If conditioning is "on", between the titrations the titration
solution is constantly maintained at the (1st) end point.
When conditioning is performed, the volume drift can
be displayed during the conditioning:
Display of drift during conditioning (ON, OFF).
Volume drift.
Type of drift correction (auto, man., OFF)
Type of drift correction: (EP – drift
∗
time)
auto: drift value at start is valid
man.: see below
OFF : no correction
Value for manual drift correction (0.0...99.9
µ
L/min)
Request of identifications after start of titration (id1,
id1&2, all, OFF)
After start, sample identifications can be requested
automatically: Only id1, id1 & id2, all three id's or no
inquiries.
Request of sample size after start of titration (value, unit,
all, OFF)
"all" the value and the unit will be requested.
Pulse output on I/O line 6 (L6, pin 1) of the remote
socket (first, all, cond., OFF)
see page 131.
Содержание Titrino 719 S
Страница 2: ......
Страница 6: ...1 2 Controls and parts 719 S Titrino 2 1 2 Controls and parts Front view of instrument...
Страница 8: ...1 2 Controls and parts 719 S Titrino 4 Rear view of instrument...
Страница 58: ......
Страница 106: ......
Страница 135: ...6 2 Pin assignment of the Remote socket 719 S Titrino 131 6 2 1 Lines of the Remote socket during the titration...