3.2 Remote control commands
719 S Titrino
85
.UnitSmpl: If "no unit" is desired for the sample size, the blank string must be
entered.
3.2.2.55. S
mplData
.ON
Silo
.D
elLine $G
S
mplData
.ON
Silo
.D
elLine
.L
ineNum 1...99,
OFF
Deletion of a silo line. The line # is deleted with &SmplData.ONSilo.DelLine
$G. If a formerly deleted line is edited again, it becomes valid (function "unde-
lete").
3.2.2.56. S
mplData
.ON
Silo
.DelA
ll $G
Deletes the entire silo memory. Must be triggered with $G.
3.2.2.57. S
mplData
.ON
Silo
.Cy
cleLines ON,
OFF
Silo data cycling.
With "ON", executed lines are copied to the next free silo lines, see page 53.
Exercise caution if you edit the silo memory during the determinations!
3.2.2.58. I
nfo
.R
eport $G
I
nfo
.R
eport
.S
elect
configuration, parameters, smpl data,
statistics, silo, calib, C-fmla, def,
user method,
full
, short, calc, all, ff
$G sends the selected report to the COM which is set in
&Config.PeriphUnit.RepToComport:
configuration: Configuration report. Is not accessible during a running deter-
mination.
parameters: Parameter report of the current method. During a running deter-
mination only "live"-parameters are accessible.
smpl data: Current sample data.
statistics: Statistics table with the individual results.
silo:
Contents of the silo memory.
calib:
Calibration data of the measuring input in the current method.
C-fmla:
Contents of the <C-fmla> key.
def:
Contents of the <def> key.
user method: Contents of the method memory.
full:
Full result report of the last completed determination.
short:
Short result report of the last completed determination.
calc:
Calculation report of the current method.
all: All
reports.
ff:
Form feed on printer.
Reports which are sent from the Titrino are marked with space (ASCII 32) and
' at the beginning. Then an individual identifier for each report follows. Reports
which are triggered by RS232 ($G) have the same introducer but without pre-
ceding space, i.e. they start with '.
3.2.2.59. I
nfo
.C
alibrationData $G
I
nfo
.C
alibrationData
.I
np1
.p
Has -20.00...
7.00
...+20.00
I
nfo
.C
alibrationData
.I
np1
.S
lope -9.999...
1.000
...+9.999
I
nfo
.C
alibrationData
.I
np1
.T
emp -170.0...
25.0
...+500.0
Содержание Titrino 719 S
Страница 2: ......
Страница 6: ...1 2 Controls and parts 719 S Titrino 2 1 2 Controls and parts Front view of instrument...
Страница 8: ...1 2 Controls and parts 719 S Titrino 4 Rear view of instrument...
Страница 58: ......
Страница 106: ......
Страница 135: ...6 2 Pin assignment of the Remote socket 719 S Titrino 131 6 2 1 Lines of the Remote socket during the titration...