3.2 Remote control commands
719 S Titrino
77
Assignment of common variables.
The values of the common variables are to be found in &Config.ComVar. They
can be viewed and entered there, see 3.2.2.50.
3.2.2.7. M
ode
.Def.R
eport
.A
ssign
SET, MEAS, CAL:
full, short, calc, param, calib, ff
Definition of the report sequence, which is outputted automatically at the end
of the determination. Entries of more than one block have to be separated with
";".
3.2.2.8. M
ode
.Def.M
ean
.1.A
ssign
RS1
, RSX, EPX, CXX
M
ode
.Def.M
ean
.2.A
ssign
etc., up to
.9
Assignment of the statistics calculations. Valid assignments are a requirement
for statistics calculations. In addition, the statistics calculation must be
switched on, see 3.2.2.22. Rules for statistics calculations see page 41.
3.2.2.9. M
ode
.C
Fmla
M
ode
.C
Fmla
.1.V
alue
0
...
±
999 999
M
ode
.C
Fmla
.2.V
alue
etc., up to
.19
Calculation constants specific to a method. Stored in the method memory of
the Titrino. Operands specific to the sample (3.2.2.52 and 3.2.2.53) and val-
ues of common variables (3.2.2.50) on the other hand are not stored with the
methods.
3.2.2.10. M
ode
.P
arameter
.S
ET1
.E
P pH:
0...
±
20.00,
OFF
U,
Ipol:
0...
±
2000,
OFF
Upol:
0...
±
200.0,
OFF
M
ode
.P
arameter
.S
ET1
.U
nitEp read
only
Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and
Ipol) resp. µA (with Upol). The corresponding unit can be read with .UnitEP.
If the value is on "OFF", no further nodes will appear from SET1.
3.2.2.11. M
ode
.P
arameter
.S
ET1
.D
yn pH:
0.01...20.00,
OFF
U,
Ipol:
1...2000,
OFF
Upol:
0.1...200.0,
OFF
M
ode
.P
arameter
.S
ET1
.UnitD
yn read
only
M
ode
.P
arameter
.S
ET1
.M
axRate 0.01...
10
...150, max.
M
ode
.P
arameter
.S
ET1
.Mi
nRate 0.01...
25.0
...9999.9
Parameters for SET: Control parameters, see page 25.
.Dyn:
Dynamics, control range in pH, mV (with U and Ipol) or µA (with
Upol). The corresponding unit can be read with .UnitDyn.
.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum
possible rate with the Exchange Unit in current use.
.MinRate: Minimum titration rate in ul/min.
Содержание Titrino 719 S
Страница 2: ......
Страница 6: ...1 2 Controls and parts 719 S Titrino 2 1 2 Controls and parts Front view of instrument...
Страница 8: ...1 2 Controls and parts 719 S Titrino 4 Rear view of instrument...
Страница 58: ......
Страница 106: ......
Страница 135: ...6 2 Pin assignment of the Remote socket 719 S Titrino 131 6 2 1 Lines of the Remote socket during the titration...