3.2 Remote control commands
719 S Titrino
65
*Parameter
Tree part "Parameters for SET"
├
.
S
ET1
Control parameters for EP1
│
├
.
E
P
Endpoint 1
depends on meas.quant. 3.2.2.10.
│
├
.
U
nitEp
Unit of endpoint
read only
ditto
│
├
.
D
yn
Dynamics
depends on meas.quant. 3.2.2.11.
│
├
.
UnitD
yn
Unit of dynamics
read only
ditto
│
├
.
M
axRate
Maximum dosing rate
0.01...
10
...150, max.
ditto
│
├
.
Mi
nRate
Minimum dosing rate
0.01...
25.0
...9999 ditto
│
├
.
S
top
Titration stop
│
│
├
.
T
ype
Type of stop criterion
drift
, time
3.2.2.12.
│
│
├
.
D
rift
Stop drift
1...
20
...999 ditto
│
│
├
.
Ti
me
Switch-off delay time
0...
10
...999, inf
ditto
│
│
├
.
S
topT
Stop time
0...999 999,
OFF
ditto
├
.
SET2
Control parameters for EP2, as for EP1
├
.
T
itrPara
Titration parameters
│
├
.
D
irection
Titration direction
+, -,
auto
3.2.2.13.
│
├
.
X
Pause
Pause 1
0
…999 999
3.2.2.14.
│
├
.
S
tart V
Start volume
│
│
├
.
T
ype
Type of start volume
abs., rel.,
OFF
3.2.2.15.
│
│
├
.
V
Volume for absolute start volume
0
...999.99 ditto
│
│
├
.
F
actor
Factor for relative start volume
0
...
±
999 999
ditto
│
│
├
.
R
ate
Dispensing rate for start volume
0.01...150.0,
max.
ditto
│
├
.
P
ause Pause
2
0
...999 999
3.2.2.16.
│
├
.
E
xtrT
Extraction time
0
...999 999
3.2.2.17.
│
├
.
M
easInput Measuring
Input
1
, 2, diff.
3.2.2.18.
│
├
.
I
pol
Polarization
current
0...
1
...
±
127 ditto
│
├
.
U
pol
Polarization voltage
0...
400
...
±
1270 ditto
│
├
.
Po
lElectrTest
Test for polarized electrodes
ON,
OFF
ditto
│
├
.
T
emp
Titration temperature
-170.0...
25.0
...500.0 3.2.2.19.
│
├
.
St
opCond
Stop conditions
│
├
.
V
Stop
Stop volume
│
│
├
.
T
ype
Type of stop volume
abs
., rel., OFF
3.2.2.20.
│
│
├
.
V
Volume for absolute stop volume
0...
99.99
...9999.99 ditto
│
│
├
.
F
actor
Factor for relative stop volume
0...
±
999 999
ditto
│
├
.
F
illRate Filling
rate
0.01...150.0,
max.
3.2.2.21.
│
├
.
Sta
tistics
Statistics
│
├
.
S
tatus
Status of statistics calculation
ON,
OFF
3.2.2.22.
│
├
.
M
eanN
No. of individual determinations
2
...20 ditto
│
├
.
R
esTab Result
table
│
│
├
.
S
elect
original
,delete n,delete all
ditto
│
│
├
.
D
elN
Deletion of individual results
1
...20 ditto
│
├
.
P
resel
Preselections
│
├
.
C
ond
Conditioning
ON,
OFF
3.2.2.23.
│
├
.
D
riftDisp
Display of drift during cond.
ON
, OFF
ditto
│
├
.
DC
or
Drift Correction
│
│
├
.
T
ype
Drift type
auto, man.,
OFF
3.2.2.24.
│
│
├
.
V
alue
Value for manual drift correction
0.0
…99.9 ditto
│
├
.
I
Req
Request of Id's after start
id1, id1&2, all,
OFF
3.2.2.25.
│
├
.
S
Req
Request of smpl size after start
value, unit, all,
OFF
ditto
│
├
.
A
ctPulse
Output of a pulse
first, all, cond.,
OFF
3.2.2.26.
Содержание Titrino 719 S
Страница 2: ......
Страница 6: ...1 2 Controls and parts 719 S Titrino 2 1 2 Controls and parts Front view of instrument...
Страница 8: ...1 2 Controls and parts 719 S Titrino 4 Rear view of instrument...
Страница 58: ......
Страница 106: ......
Страница 135: ...6 2 Pin assignment of the Remote socket 719 S Titrino 131 6 2 1 Lines of the Remote socket during the titration...