![Metrohm Titrino 719 S Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/metrohm/titrino-719-s/titrino-719-s_instructions-for-use-manual_1779645017.webp)
2.4
Tutorial
719 S Titrino
13
<ENTER>
<SELECT>
<ENTER>
2 x <QUIT>
Calculate the content of your hydrochloric acid in g/L:
RS1=EP1*C01*C02/C00
End point*conc.(titrant)*molecular weight/sample size
To correct a formula, delete it with <CLEAR>.
•
Confirm the formula with <ENTER>.
•
You may enter a text for the result output, see page
8.
•
Enter the desired number of decimal places for the
result.
•
Select the desired unit g/L with <SELECT>, or type
a text as unit, see page 8.
The Titrino prompts for the calculation of the next result.
Quit the formula entry by pressing <QUIT> twice.
In order to be able to calculate the result, enter the calculation constants used in the for-
mula.
Entry of the calculation constants
<C-FMLA>
0.1
<ENTER>
36.47
<ENTER>
•
Press <C-FMLA>.
The constants which have been used in the formula
are requested:
input with digit keys, confirm with
<ENTER>.
C01: Concentration of your titrant= 0.1 mol/L.
Enter 0.1.
C02: Molar mass of HCl = 36.47 g/mol
Enter 36.47.
Now only the constant C00 is missing to perform the calculation. This constant is re-
served for the sample size. It can be entered directly.
RS1=EP1*C01*C02/C00
>formula
RS1 text
RS1
>formula
RS1 decimal places
2
>formula
RS1 unit
%
RS?
C-fmla
>C01 0.0
C-fmla
>C02 0.0
Содержание Titrino 719 S
Страница 2: ......
Страница 6: ...1 2 Controls and parts 719 S Titrino 2 1 2 Controls and parts Front view of instrument...
Страница 8: ...1 2 Controls and parts 719 S Titrino 4 Rear view of instrument...
Страница 58: ......
Страница 106: ......
Страница 135: ...6 2 Pin assignment of the Remote socket 719 S Titrino 131 6 2 1 Lines of the Remote socket during the titration...