96
mains test socket.
*Delay time
Delay [0.2 s … 5 s]
Test limits
H Limit (Idiff)
H limit [Off, Custom, 0.25 mA ... 15.0 mA ]
L Limit (Idiff)
L limit [Off, Custom, 0.25 mA ... 15.0 mA ]
Output connections
[Socket L,N
– PE,P/S ]
Test circuit
Figure 6.47: Measurement of Differential Leakage current
Differential Leakage measurement procedure
Select the
Differential Leakage
function.
Set test parameters / limits.
Connect device under test to mains test socket and optionally to P/S terminal.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 6.48: Examples of Differential Leakage measurement results