MI 3394 CE
MultiTester
XA
Single tests
90
Test circuit
Figure 6.34: HV DC programmable test
HV DC programmable test procedure
Select the
HV DC
programmable
function.
Set test parameters / limits.
Connect HV test leads to HV(~,+) and HV(~,-) terminals on the instrument.
Connect HV test leads to device under test.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 6.35: Examples of HV DC programmable test results
Note:
First HV measurement after power on the instrument (if password protection is enabled)
or first HV measurement after enabling or changing password require entering password
for enabling HV test. Refer to chapter
4.6.7 Change password for HV functions
information.