MI 3394 CE
MultiTester
XA
Auto Sequences
®
116
on
Sets operation mode for multiple points. For more information
see chapter
7.2.1.5 Managing multiple points
7.2.1.3
Auto Sequence
®
Configurator menu
Auto Sequence® Configurator menu options are active only when single tests within selected
Auto sequence® contain configurable limits and / or parameters. They can be changed to suit
requirements of current object under test before Auto sequence® execution. Original settings
are overwritten for intended Auto sequence® execution only.
Select
Configurator option from Auto sequence® Main screen or View screen to open
Configurator menu
presented on
below
.
Available settings are organized in groups, each group starts with concerned Single test name.
Limit Calculator is referred to Continuity function. Refer to Single test description chapter for
details of parameters and limits setting / calculation.
Figure 7.6: Auto Sequence
®
Configurator menu
– opened from Auto Sequence
®
Main menu on the
left, on the right opened from Auto Sequence
®
view menu
Options
Starts
the
Auto
Sequence®
from
Configurator
menu.
Option is available, if configurator was opened from Auto Sequence®
Main menu. New configuration is automatically applied to all related
single tests.
Confirms limits and parameters setting and returns to view menu.
Option is available, if configurator was opened from Auto Sequence®
view menu. Start the Auto Sequence® from view menu with confirmed
configuration.
Содержание MI 3394 CE MultiTesterXA
Страница 1: ...CE MultiTesterXA MI 3394 Instruction manual Ver 3 4 13 Code no 20 752 432...
Страница 80: ...MI 3394 CE MultiTesterXA Single tests 79 Figure 6 19 Examples of Continuity measurement results...
Страница 103: ...MI 3394 CE MultiTesterXA Single tests 102 Figure 6 57 Examples of Power measurement results...
Страница 110: ...MI 3394 CE MultiTesterXA Single tests 109 Save results optional Figure 6 69 Examples of Functional Inspection results...