MI 3394 CE
MultiTester
XA
Technical specifications
128
Test terminals
Function
Connections
Withstanding voltage (HV
DC
, HV
DC-P
)
HV(
~,+) ↔ HV(~,-)
10.3 Continuity
Continuity
Range
Resolution
Accuracy
R
0.00
... 19.99
0.01
(2 % of r 2 D)
20.0
... 99.9
0.1
3 % of reading
100.0
... 199.9
0.1
5 % of reading
200
... 999
1
indicative
Voltage drop ( Iout = 10 A)
Range
Resolution
Accuracy
ΔU
0.00 V ... 19.99 V
0.01 V
(2 % of r 5 D)
20.0 V ... 99.9 V
0.1 V
3 % of reading
Limit value of voltage drop versus wire cross-section:
Wire cross-section (mm
2
)
Limit voltage drop (V)
0.5
5.0
0.75
5.0
1
3.3
1.5
2.6
2.5
1.9
4
1.4
6
1.0
Operating range (acc. to EN 61557-4) .............
0.08 Ω ... 199.9
Test currents ................................................... 0.2A, 4A, 10A, 25A
Current source (at nominal mains voltage, use of standard accessories)
........................................................................
0.2 A at R < 8 Ω
........................................................................
4 A at R < 1 Ω
........................................................................ 10 A
at R < 0.5 Ω
........................................................................
25 A at R < 0.2 Ω
Open circuit voltage ......................................... < 6 V a.c.
Max current lead resistance ............................
40 Ω
Test lead compensation (P/S
– PE) ................ up to 5 Ω
Test method: Continuity 4 wire ....................... Kelvin method, floating to earth
Test method: Continuity P/S
– PE.................... 2 wire test, floating to earth
Test terminals
Function
Connections
Continuity P/S
–
PE
P/S ↔ Mains test socket (PE), TC1
Continuity 4 wire
P1/C1 ↔ P2/C2
Содержание MI 3394 CE MultiTesterXA
Страница 1: ...CE MultiTesterXA MI 3394 Instruction manual Ver 3 4 13 Code no 20 752 432...
Страница 80: ...MI 3394 CE MultiTesterXA Single tests 79 Figure 6 19 Examples of Continuity measurement results...
Страница 103: ...MI 3394 CE MultiTesterXA Single tests 102 Figure 6 57 Examples of Power measurement results...
Страница 110: ...MI 3394 CE MultiTesterXA Single tests 109 Save results optional Figure 6 69 Examples of Functional Inspection results...