MI 3394 CE
MultiTester
XA
Single tests
107
Figure 6.64: Examples of Discharging Time measurement results (Output = External)
Measuring principle (Output = Socket)
The measuring principle of the Discharging time function is as following:
Phase
The DEVICE UNDER TEST is connected to the mains test socket.The instrument
monitors the mains voltage and internally stores the peak voltage value.
Phase
The instrument disconnects the DEVICE UNDER TEST from the supply and the
voltage at the supply connections starts to fall. Disconnection moment is always at
peak voltage.
Phase
After the voltage drops below the limit value the timer is stopped.
Test circuit (Output = Socket)
Figure 6.65: Discharging Time test (Output = Socket)
Discharging Time test procedure (Output = Socket)
Select the
Discharging Time
function.
Set test parameters / limits.
Connect the device under test to the mains test socket on the instrument.
Start measurement.
Measurement can be stopped manually or automatically.
Save results (optional).