E.3 Elements of an Auto Sequence
®
E.3.1
Auto Sequence
®
steps
There are three kinds of Auto Sequence
®
steps.
Header
The Header step is empty by default.
Flow commands can be added to the Header step.
Measurement step
The Measurement step contains a Single test and the Operation after end of test flow command
by default. Other Flow commands can also be added to the Measurement step.
Result
The Result step contains the Result screen flow command by default. Other Flow commands
can also be added to the Result step.
E.3.2
Single tests
Single tests are the same as in Metrel ES Manager Measurement menu.
Limits and parameters of the measurements can be set. Results and sub-
results can’t be set.
E.3.3
Flow commands
Flow commands are used to control the flow of measurements. Refer to chapter
E.3.4
Number of measurement steps
Often the same measurement step has to be performed on multiple points on the device under
test. It is possible to set how many times a Measurement step will be repeated. All carried out
individual Single test results are stored in the Auto Sequence
®
result as if they were
programmed as independent measuring steps.
Содержание MI 3394 CE MultiTesterXA
Страница 1: ...CE MultiTesterXA MI 3394 Instruction manual Ver 3 4 13 Code no 20 752 432...
Страница 80: ...MI 3394 CE MultiTesterXA Single tests 79 Figure 6 19 Examples of Continuity measurement results...
Страница 103: ...MI 3394 CE MultiTesterXA Single tests 102 Figure 6 57 Examples of Power measurement results...
Страница 110: ...MI 3394 CE MultiTesterXA Single tests 109 Save results optional Figure 6 69 Examples of Functional Inspection results...