Malfunctions, warning and error messages
Error description
Cause of error
Troubleshooting
Error message AI E-3 in the temperature
display
TEMP
AI E-3
Set 37.0
°C
37
.4
°C
Temperature monitoring sensor is de-
fective. The operating sensor takes over
the measurement function.
■
The appliance can temporarily be kept
in service
■
Contact customer service as soon as
possible
Error message E-3 in the temperature
display
TEMP
Set 45.0
°C
E-3
°C
Operating and monitoring sensor defective
■
Switch off appliance
■
Remove the chamber load
■
Contact customer service
Error message E-6 in the humidity
display
HUMIDITY
Set 50
.0%rh
%rh
E-6
Humidity sensor defective
■
No humidity control possible
■
Contact customer service
When switching on the appliance, the
start animation is displayed in another
colour than white
Cyan
:
Not enough storage space on the SD
card
■
Contact customer service
Red
:
The system
fi
les could not be loaded
Orange
:
The fonts and images could not be load-
ed
6.2 Power
failure
In case of a power failure, the appliance operates as follows:
In manual mode
After power supply has been restored, operation is continued with the parameters set.
The time and the duration of the power failure is documented in the log memory.
In timer or program mode
In case of an interruption of the power supply of less than 60 minutes, the current pro-
gramme is continued from the point at which it was interrupted. For longer interruptions
of the power supply, all appliance functions (heating, fan etc.) are switched off.
In remote control mode
The previous values are restored. If a program has been initiated via remote control, it
is continued.