57
03/2021
Menu mode
7.8 Protocol
The appliance continually logs all relevant measured values, settings and error messag-
es at 1-minute intervals. The internal log memory is of the continuous memory type. The
logging function cannot be switched off and is always active. The measured data are
stored in the appliance, safe from manipulation. If the power supply is interrupted, the
time of the power failure and voltage recovery are stored in the appliance.
You can read out the log data for different periods to a USB storage medium via the USB
port or via Ethernet and then import them to the AtmoCONTROL program for graphical
representation, printing out or storage.
The log memory of the appliance is not modi
fi
ed or deleted by reading it out.
1. Connect the USB storage medium to the USB port on the right of the
ControlCOCKPIT.
Protocol
Last download 12.10.2012 12:00
1 month
Complete control range
1 week
2. Activate the protocol. To do so, press the activation key on the right side of the
PROTOCOL display. The display is enlarged and the period This month automati-
cally highlighted. To select another logging period, use the turn control.
Protocol
1 Month
Complete control range
1 Week
3. Save your selection by pressing the con
fi
rmation key. The transfer starts and a
status symbol indicates the progress.
Protocol
l
D
l d 25 10 2012 13 20
1 Month
Complete control range
1 Week
4. As soon as the transfer is complete, a check mark appears in front of the period
selected. You can now remove the USB storage medium.
For a description of how to import and process protocol data in AtmoCONTROL or read
it out via Ethernet, please refer to the separate AtmoCONTROL manual.
i