30
03/2021
Operation and control
5.3.1 Manual
mode
In this operating mode, the appliance runs in permanent operation at the values set on
the ControlCOCKPIT.
Adjustment options
As described in chapter 5.2.2 , you can set the following parameters after pressing the
corresponding activation key (in any sequence):
Temperature
°C
22
.4
TEMP
°C
Set 37.0
Adjustment range depends on appliance (see name plate chapter 2.7 and technical
data chapter 2.8)
Heating operation is indicated by the symbol.
Cooling is indicated by the symbol.
You can select °C or °F as the temperature unit displayed.
i
The minimum temperature that can be reached depends on the surrounding condi-
tions. The devices can cool down to 20 °C below room temperature. For this purpose,
the Peltier module needs suf
fi
cient ventilation
Humidity (for constant climate chambers HPPeco only)
HUMIDITY
70
%rh
Set
34
% rh
Adjustment range: 10 to 90 % rh
Humidi
fi
cation is indicated by the symbol.
Dehumidi
fi
cation is indicated by the symbol.
i
A high level of air humidity in the interior can only be achieved without condensa-
tion if the interior is thoroughly heated. For this reason, how fast the humidity is
dynamically adjusted to approach the setpoint depends on the interior temperature.
Interior lighting (only for models with light module)
LIGHT
%
0
Temp too high
LIGHT
%
40
Adjustment range: 0 to 100 % in steps of 1%
i
The interior light is only active at temperatures of up to 40°. If this temperature is
exceeded, the interior light switches off automatically. The light display will then
show “Temp too high”.