22
03/2021
Putting into operation
4.
Putting into operation
Condensation in the electrical components may cause short circuits. After transport-
ing or storing the device under humid conditions, remove it from its packaging and let
it ventilate for at least 24 hours in normal environmental conditions. Do not connect
the device to the mains power during this time.
When putting the appliance into operation for the
fi
rst time, do not leave it unattended
until it has reached a steady state.
4.1
Connecting the appliance
Observe the country-speci
fi
c regulations when making connections (e.g. DIN VDE
0100 with earth leakage circuit breaker, in Germany). Observe the connection and pow-
er ratings (see name plate and the "Technical Data"chapters 2.7 and 2.8). Make sure to
establish a safe PE conductor connection.
Plug the provided power cable into the rear of the appliance and connect it to the
power supply.
Place the power cable so that
■
it is easily accessible at all times and can be pulled off quickly, for example in case of
interference or an emergency
■
it does not represent a trip hazard
■
it cannot come into contact with any hot parts
WARNING