Menu mode
7.3.5 Slide-in unit type steel grid or shelf
Type of slide-in unit (steel grid or shelf)
Here, you have to select the type of slide-in unit (steel grid or shelf) used. The selection
Shelf enables you to adjust the control function to the different air
fl
ow characteristics
in the interior when using optional sliding shelves instead of the steel grids that are
provided as standard.
7.3.6 Balance
For appliances of the sizes 260, 410 and 750, application-speci
fi
c correction of the heat
output distribution (balance) between the upper and lower heating groups is possible.
The adjustment range is from –50 % to +50 %.
Distribution of the heating/cooling power (example):
The –20 % (left) setting causes the lower Peltier elements to work at 20 % less power
than the upper ones. The +30 % (right) setting causes the lower Peltier elements to
work at 30 % more power than the upper ones. The 0 % setting restores the default
distribution settings.
7.3.7
Dehumidi
fi
cation interval
The dehumidi
fi
cation peltier modules behind the rear panel precisely generate cold
spots inside the chamber in order to remove humidity from the appliance in a controlled
way.
If the device is dehumidifying for a long period of time in the lower end of the climate
diagram, the water in the air will freeze at the dehumidi
fi
cation peltier modules. If solid
ice should form at the rear panel around the dehumidi
fi
cation peltier modules, the dehu-
midi
fi
cation interval must be adjusted.
SETUP
Unit
IP adress
255.145.136.225
Subnet mask
255.255.0.0
°C
°F
Timer mode
Alarm temp
TWW
TWB
Slide-in unit
Grid Shelf
upper heat output
upper heat output
lower heat output
-20%
lower heat output
+30%
SETUP
Balance
-20
%
2/2
SETUP
Balance
+30
%
2/2