54
03/2021
Menu mode
Example: Humidity deviation at 60 % should be corrected
%rh
%rh
%
0,5
-
%
Temperature
Cal1
Cal2
40.0
Humidity
60.0
1,0
+
%rh
%
1,0
+
Cal3
80.0
5. Save the setting by pressing the con
fi
rmation key. The corresponding calibration
value is automatically highlighted.
%rh
%rh
%
0,5
-
%
Temperature
Cal1
Cal2
40.0
Humidity
60.0
0,0
%rh
%
1,0
+
Cal3
80.0
6. Set the calibration value to 0.0
% and accept the setting by pressing the con
fi
r-
mation key.
HUMIDITY
29
.5%rh
Set
60
.0
%rh
7. Position the sensor of the calibrated reference instrument centrally in the working
chamber of the appliance.
8. Close the door and, in manual mode, adjust the set humidity to 60 % rh.
58.5
%rh
9. Wait until the appliance reaches the set humidity and displays 60 % rh. The refer-
ence instrument displays for example 58.5 %
rh.
%rh
%rh
%
0,5
-
%
Temperature
letzte Kalibrierung 12 10 2012 12 00
Cal1
Cal2
40.0
Humidity
60.0
-1.5
%rh
%
1,0
+
Cal3
80.0
10. In the SETUP, adjust the compensation correction value Cal2 to –1.5% (actual
value measured minus setpoint temperature) and save the setting by pressing the
con
fi
rmation key.
60.0
%rh
HUMIDITY
60
.0%rh
Set 60
.0%rh
11. After the calibration-, the humidity measured by the reference instrument should
now also be 60 % rh.