52
03/2021
Menu mode
Example: Temperature deviation at 30°C is to be corrected
31.6
°C
30
.0°C
TEMP
Set 30
.0°C
8. Wait until the appliance reaches the set temperature and displays 30
°C. The ref-
erence instrument displays for example 31.6
°C.
C
K
0,2
-
C
K
C
K
0,2
-
Temperature
Cal1
Cal2
Cal3
40.0
5.0
30.0
1,6
+
9. In the SETUP, adjust the calibration value Cal2 to +1.6 K (actual value measured
minus setpoint temperature) and save the setting by pressing the con
fi
rmation
key.
30,0
°C
30
.0°C
TEMP
Set 30
.0°C
10. After the calibration procedure, the temperature measured by the reference in-
strument should now also be 30
°C.
With Cal1, a calibration temperature below Cal2 can be programmed accordingly, and
with Cal3, a temperature above. The minimum difference between the Cal values is 10
K.
If all calibration values are set to 0.0 K, the factory calibration settings are restored.