Malfunctions, warning and error messages
6.
Malfunctions, warning and error messages
Live components may be exposed once the covers have been removed. Touching
these can lead to an electric shock. Malfunctions requiring work inside the appliance
may only be recti
fi
ed by electricians. Observe the separate service manual for this.
Do not try to rectify appliance errors yourself but contact the MEMMERT customer ser-
vice department or an authorised service point.
In case of enquiries, please always specify the model and appliance number on the
nameplate (see chapter 2.7).
6.1
Warning message of the monitoring function
If the alarm sound has been activated in the menu mode ( sound, chapter 7.7, indicated
by the speaker symbol ), the alarm is additionally signalled by an intermittent acous-
tic signal. If the con
fi
rmation key is pressed, the acoustic alarm can be temporarily
switched off until the next alarm event occurs.
6.1.1 Temperature
monitoring
Description
Cause
Action
Temperature alarm and "ASF" are
displayed
TEMP
ASF Set 38.5
°C
40
.4
°C
Automatic temperature monitor (ASF)
was triggered.
■
Check if the door is closed. Closing the
door
■
Extend the ASF tolerance band
■
If the alarm continues: Contact custom-
er service
Temperature alarm and "TWW" are
displayed
TEMP
TWW Set 38.5
°C
42
.4
°C
The adjustable temperature controller
(TWW) has assumed heating control.
■
Increase the difference between the
monitoring and setpoint temperature –
by either increasing the max value of the
temperature monitoring or decreasing
the setpoint temperature.
■
If the alarm continues:
Contact customer service
WARNING