background image

MODEL 6540-1 

INST-PM-6540-1-UVA 

Содержание 6540-1

Страница 1: ...MODEL 6540 1 INST PM 6540 1 UVA ...

Страница 2: ...Revision Log Revisions Description Letter Date A 12 06 2016 Initial release B 04 04 2018 New cover ...

Страница 3: ...el 6540 1 are shown below Note Lamps may be black or white depending on kit Revision Level Note Photos may have a different appearance depending on chamber options 1 Turn off chamber and unplug power cord INSTRUCTIONS FOR REPLACING UVA LAMPS IN MODEL 6540 1 REPLACING LAMPS Disconnect electrical power before removing light banks Lamps ...

Страница 4: ... light bank shelf and place lamp side up on flat surface as shown below 4 Remove and discard old lamp by rotating lamp turn and lifting out of lamp sockets Repeat for each lamp to be replaced 5 Insert new lamp ends into lamp sockets and rotate turn Repeat for each lamp to be replaced Top Light Bank Shelf ...

Страница 5: ...hamber and press firmly to seat rear connector as shown below 7 Plug in power cord and turn on chamber 8 Reset lamp life hours by pressing Setup Menu button on right side of MAIN SCREEN as shown below Light Bank Shelf Setup Menu Button Press Here to Seat Rear Connector ...

Страница 6: ...Rev B 04 04 2018 9 Press LAMP LIFE button as shown below 10 Press RESET UVA LAMP HOURS button as shown below 11 Press Check Mark button as shown below RESET UVA LAMP HOURS Button Check Mark Button Button LAMP LIFE Button ...

Страница 7: ... 6540 1 UVA Rev B 04 04 2018 12 Press Check Mark button to confirm selection as shown below 13 Press Check Mark button to return to MAIN SCREEN as shown below Check Mark Button Button Check Mark Button Button ...

Отзывы:

Похожие инструкции для 6540-1