3.8.3.5 Overcurrent Relay Instantaneous Test Results
This screen displays the results upon test completion. No data may be entered in this screen.
Figure 46 Overcurrent Relay Instantaneous Test Screen
The Timing, Target & Seal-In, and Pickup test screens may be entered to change test settings.
Each current channel must be selected by touching I1, I2, and I3. However, there must be a
corresponding V/I Generator.
The bottom of the screen has common buttons.
File:
This selection will take the user to the File
Manager.
Batt Sim Off / Batt Sim On:
This selection will turn the Battery Simulator On
and Off.
Test Menu:
This selection will return the controller to the
Test Menu Screen.
Main Menu:
This selection will return the controller to the
Main Menu
Start / Stop:
This selection will apply the set quantities
displayed and begin the test. After Start is
pressed this button will turn red and display
Stop. Pressing Stop at any time will stop the
test and turn the outputs off.
A link to Timer Setup configuration is be available on this screen. Also the Timer Setup
configuration is displayed for convenience. The user selects the appropriate number of channels
required pressing I
1
,I
2
and I
3
.The amount of current required will be split proportionally between
selected current channels. The status of the relay contact will be displayed on the screen real
time along with the value of the current(s) and the percentage error.
68
www
. ElectricalPartManuals
. com