3.8.10.4
Manual Transducer Test Settings Screen
Figure 98 Transducer Setting Screen
This screen is used for selection of single phase or three-phase transducers such as; AC and DC
Voltage, AC and DC Current, Frequency, Power (Watts), Reactive Power (VAR's), Apparent
Power (VA) and Power Factor. The types of transducers are easily selected in the
Transducer
Setting Screen
.
3.8.10.4.1 Select Transducer Type
In
the
Transducer Setting Screen
, the user may select from a variety of transducers,
see
Select Transducer Type
window. Here the operator may select, by touching the
selection window, the type of transducer that he/she needs to test. Watt, VAR and VA
transducers come in 1, 1 1/2, 2 , 2 1/2 and 3 element configurations. Selecting the
number of elements will automatically select the appropriate number of output voltages
and currents needed to test the selected transducer. For example, selection of a single
element Watt transducer, V1 and I1 sources will automatically be selected for you. In the
case of a three-phase 3 element transducer V1,V2, V3, I1, I2 and I3 will be preselected
for your use. Selections available are:
Single Phase
Multi-Phase
AC Volts
Watts / VAR/ VA - 1 1/2 Element
AC Current
Watts / VAR / VA - 2 Element
DC Volts
Watts / VAR / VA - 2 1/2 Element
DC Current
Watts / VAR / VA / Power Factor - 3 Element
Frequency
Watts / VAR / VA / Power Factor - 1 Element
3.8.10.4.2
Transducer Description Section
In the description windows the operator enters descriptive information relative to the
transducer to be tested. This information will be saved with the test results. The following
describes the window entries.
-
135