Multi-Directional Nozzles
The nozzles also need little maintenance, the nozzles should stay clean in the position they are in. However, the pins
which give out the ionisation will need to be monitored.
The pins are made from titanium so are very hard wearing but they may need to be cleaned periodically (every 3
months) The 977CM will alarm when the pins are becoming contaminated.
To clean the pins a tooth brush can be used, in high contamination or degradation cases the pins can be removed
and replaced.
Vacuum Compartment
The Cleaning area should be cleaned once a week, whilst there is a flow of air through the cabinet there are still
dead areas where dirt and contamination can build up. This can be cleaned out with a microfibre cloth.
The vacuum perforated plate should also be removed when the cleaning takes place.
When the plate is removed any contamination that is on the vacuum plate can be removed or swept into the filter
bag.
Filter Compartment
In the filter compartment, there should be little to no contamination due to the Bag filter being directly below the
vacuum compartment collecting all of the contamination coming off the components being cleaned.
However, the filter compartment can still be cleaned out using a cloth, the HEPA and bag filter will need to be
removed before the cleaning takes place. When cleaning care must be taken so no contamination falls into the 6
pumps as this can cause damage.