8.1 EMC guidance and manufacturer's declaration
Guidance and manufacturer’s declaration – electromagnetic emissions
The TELESMART holter recorder is intended for use in the electromagnetic environment
specified below. The customer or the user of the TELESMART holter recorder should
assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The TELESMART holter recorder uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The TELESMART holter recorder is suitable for
use in all establishments other than domestic
and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The TELESMART holter recorder is intended for use in the electromagnetic environment specified below.
The customer or the user of the TELESMART holter recorder should assure that it is used in such an
environment.
IMMUNITY test
IEC 60601 test level
Compliance level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
TELESMART-H/P - Holter-ECG-Recording System - version D