Operation
Page 23
existant?
Back
No
Yes
The [back] button returns to the entry menu, the [yes] and [no] buttons lead to
system controls:
The following appears:
Back
Start
[A]
[Back] returns to the entry menu. The [[A]
► ►
] button allows the user to switch
between channels. [Start] begins the recording.
When the recording has started, the following appears:
Recording running
Duration:
05:24:23
Marker
Pressing the [Marker] button places a marker in the recording after 300 ms, and is
confirmed with two beeps. A dead-time is maintained between two markers where
no additional markers can be added.
The recording will be stopped after the specified time, if a low voltage condition ex-
ists, if the memory card is full, or if a user interrupt is signaled.
4.3.2 Starting the recorder with PADSY
When the recorder is turned on, the entry menu is displayed:
Transferred recording
Last name, first
name
Duration: 25:24
New
PADSY
To connect the recorder to the PC, press the [PADSY] button.
The following display appears:
Connecting with
PADSY
If the recorder finds multiple computers, a list of the detected computers is
displayed:
TELESMART-H/P - Holter-ECG-Recording System - version D