Operator’s Manual
36
IX. CLEANING AND DISINFECTION
SURFACE CLEANING
After use, the external surfaces of the Arctic Sun™
Temperature Management System may be a potential
biohazard. The Arctic Sun™ Temperature Management
System external surfaces should be manually cleaned,
disinfected and visually examined per the instructions below.
Prior to cleaning
:
1) Use appropriate personal protective equipment (e.g.
gloves, eyewear, face mask or shields) per manufacturer’s
guidelines for the enzymatic spray to protect user from
exposure to both chemicals and microorganisms.
2) Apply the castor brakes on the Arctic Sun™ Temperature
Management System.
3) Unplug the power cable from the wall outlet.
4) Straighten the power cable.
5) Disconnect all other cables and hoses from the Arctic
Sun™ Temperature Management System and straighten
them.
Manual Cleaning:
•
Timing
Clean the Arctic Sun™ Temperature Management System
external surfaces as soon as practical after use (e.g. at the
point of use).
•
Cleaning Materials
Required: Neutral-pH Enzymatic spray cleaner; clean, dry
cloths
Validated
: Medline Bio-Zolve Instrument Presoak Spray
•
Cleaning
1) Saturate a clean, dry cloth with the Enzymatic cleaner
and remove all heavy soil loads from the external
surfaces of the Arctic Sun™ Temperature
Management System.
2) Saturate a second clean, dry cloth with the Enzymatic
cleaner and thoroughly wipe all external surfaces of
the device.
a) Ensure that all surfaces are dampened,
including seams of the device.
b) Use as many additional clean cloths
saturated with the Enzymatic cleaner as
necessary to ensure device is completely
dampened.
c) Thoroughly wipe the following with additional
clean cloths saturated with the Enzymatic
cleaner – fluid delivery line, power cable,
temperature cables, USB cable, RS232
cable.
3) Allow surfaces to remain treated for a minimum of two
(2) minutes.
4) Use a clean, dry cloth to remove remaining cleaning
solution.
SURFACE DISINFECTION
Disinfect the Arctic Sun™ Temperature Management System
immediately following cleaning. Follow all manufacturers’
guidelines for the pre-saturated germicidal wipes, including
wearing gloves and other personal protective equipment when
dispensing and using the wipes.
Manual Disinfection:
•
Disinfection Materials:
Required: Pre-saturated germicidal wipes (active
ingredient – alcohol)
Validated
: Super Sani-Cloth Germicidal Disposable Wipe
•
Disinfection
1) Using pre-saturated germicidal wipes, disinfect all
external surfaces of the Arctic Sun™ Temperature
Management System.
a) Ensure that all surfaces are dampened,
including all seams of the device.
b) Use as many clean towelettes as necessary
to ensure device is completely dampened
with disinfectant.
c) Thoroughly wipe the following with additional
clean towelettes – fluid delivery line, power
cable, temperature cables, USB cable,
RS232 cable.
2) Allow the treated surfaces to remain wet for the
manufacturer’s specified contact time.
3) Wipe each surface of the device with a lint-free cloth
premoistened with sterile water.
4) Let air dry.
5) Reattach cables and hoses.
VISUAL INSPECTION AND RETURN TO USE
1) After cleaning and disinfecting the device per the
instruction provided, examine the device for cleanliness.
If visible soil remains, repeat manual cleaning
instructions.
2) Inspect cables and hoses for signs of damage.
3) Inspect external shell for discoloration or cracking.
4) Inspect labels for legibility.
5) Do not use the device if it has failed visual inspection for
soil after multiple cleaning attempts, or if there is visible
damage. Contact BD Customer Support for additional
recommendations.
6) Store the device in a dry and clean environment when not
in use.
NOTE:
The device has been validated for exterior surface
cleaning and disinfection per the instructions provided. Any
deviation from the recommended parameters must be
validated by the user. Interior surface of the device (including
deep crevices) are not addressed as part of this cleaning and
disinfection protocol.
To prevent possible discoloration, do not use iodine-based
solutions, such as Betadine, on any part of the machine.
For questions on device cleaning and disinfection of external
surfaces contact BD Customer Support.
Содержание BD Arctic Sun 6000
Страница 1: ...14868...
Страница 2: ...Operator s Manual 2 tube...
Страница 57: ...The current Operator s Manual can be found at www medivance com manuals English 57 PK2800300 MM YYY...