PATIENT TEMPERATURE (Continued)
•
Verify patient temperature
◦
The patient temperature probe is properly placed and
providing an accurate and stable temperature.
◦
Confirm patient’s temperature with a second device to
ensure accuracy.
•
Verify the custom parameters.
◦
Patient target is set to the correct temperature.
See
VI. Operation Guide–Therapy Settings
on
setting patient target temperature.
◦
One of the automatic patient control modes (e.g.
Control Patient, Cooling or Rewarming) is activated.
(The system is not in Manual Control mode.)
◦
The high water temperature limit and low water
temperature limits are set correctly. See
VI.
Operation Guide–Therapy Settings
.
•
Verify medications / shivering control
◦
If not sufficiently sedated, a patient may generate
significant heat from shivering, which will interfere
with patient temperature control. Consider
administration of additional medication for shivering
control, adequate for the patient weight and
magnitude of shivering and evaluate patient response
to medication.
◦
Has the patiently received a sedatives or other
vasoactive drugs? These drugs may cause
vasodilation, resulting in cold or warm peripheral
blood returning quickly to the core. This may lead to a
rapid change in core temperature that the system is
unable to counteract immediately.
No Patient Temperature Display
•
Confirm that the patient temperature probe is properly
placed in the patient.
•
Confirm that the patient temperature probe is connected to
the Temp In 1 cable / connector on the back of the Control
Module.
•
Confirm that the connection between the temperature
probe and cable is secure.
•
Confirm that the connection between the temperature
cable and Control Module is secure.
•
Confirm that none of the connections are wet.
•
If using a Foley temperature probe, check for adequate
urine flow.
SCREEN
Screen Locked
After approximately 2 minutes of inactivity, the screen will lock
to prevent inadvertently changing parameters during patient
therapy.
A therapy status screen will appear over the
Therapy
screen
when the screen is locked.
1. To unlock the therapy status screen and return to the
Hypothermia or Normothermia Therapy
screen,
press the button
.
2. To manually lock the screen when treating a patient,
press the
Screen Lock
button (
).
Note: The
Screen lock
button is not available when the device
is in
Stop
mode.
WIRELESS CONNECTION
Wireless Connection Lost
◦
Confirm the connectivity icon is dark blue with a slash
through it as shown in
Figure VI-3
◦
See
VI. Operation Guide–Advanced Setup
on
Connectivity Settings for instructions.
◦
If network is not connecting, confirm that network
name and password are correct
◦
To test the server, select
Start
under the Test
Connection tab